Available languages

  • ENGLISH, page 26
Safety instructions
12K42
12K44
12K50
ErgoArm plus
ErgoArm Hybrid plus
ErgoArm Electronic plus
Note: For proper functioning of the AFB
flexion compensation mechanism, the band
cam must reach its extension stop when
the elbow is fully extended.

Control harness cable attachment

The bowing harness should be attached in
the marked area.
Important!
Take care during the attachment so as to
avoid the penetration of the control harness
cable into the socket interior and possible
damaging of system components.
We recommend following one of the two
1
fixing examples seen in the pictures above.
When using the clamp stopple, use the
21A207 Clamp Stopple Set and proceed
according to the 647G231 Instructions
for Use.
21A207
12K42 ErgoArm plus, 12K44 ErgoArm Hybrid plus, 12K50 ErgoArm Electronic plus
Ottobock | 45