TROUBLESHOOTING
1.
Stall Behavior
•
If the HPQA stalls (motor trying to turn but no physical motion) the following steps should
be taken
•
Press the stop button on the LCD to disable the motor
•
Clear the obstruction that is causing the movement to be interrupted
•
Re-home the instrument by hitting the home button on the LCD
2.
Bent Spindle
•
If a spindle becomes bent this indicates that the spindle is striking the bottom of the
sample container during a test. To prevent this issue the following should be observed
•
When setting the lower position of the test ensure that the stop completely
depresses the limit switch (silver button on sled) prior to contacting the bottom of the
sample container
•
Ensure the screw on the stop is tight
3.
Tip Warning
•
If an Error 105 occurs on the screen the instrument has tilted past 7.5 degrees off of
vertical. To resolve this issue perform the following
•
Turn off HPQA from power switch on back
•
Clear obstruction that caused the tip condition. Check cabling and the spindle for
contact with the bottom of the sample container
•
Once obstructions are clear power HPQA on
•
Re-home the unit
M-HPQA-001
H E L I PAT H Q U I C K A C T I O N L A B O R AT O R Y S TA N D - O P E R AT I O N A L M A N U A L