4.
False Tip Warning
•
If an Error 105 occurs and the instrument is not titled past 7.5 degree a false detection
has occurred. To resolve this issue check the following:
•
Vibration of the instrument is the most likely cause of a false tip detection. Ensure
the HPQA is on a stable platform
5.
Instrument becomes "stuck" at the lower limit switch
•
If the instrument stops during a test and becomes "stuck" at the lower limit switch
perform the following
•
Power cycle the unit by turning power off and on from the switch on the back of the
HPQA
•
On power-up press the home button
6.
Instrument becomes stuck at the upper limit switch
•
If the instrument stops at any time and becomes "stuck" at the top limit switch
•
Turn off power to the unit from the switch on the back of the HPQA
•
Remove the power cable from the back of the HPQA
•
Physically push the sled assembly away from the top limit switch. If prior two step are
not completely you will not be able to move the HPQA head assembly
•
Re-connect the power cable and turn ON the HPQA
•
On power-up press the home button
7.
Suspected loss of position
•
If the user suspects that the home position is slipping or has moved from the original
position perform the following
•
Pres the home button on the LCD and allow the unit to perform a new homing
operation. This will reset the home position