Ametek BROOKFIELD HPQA Series Operation Instructions Manual page 31

Table of Contents

Advertisement

4.
False Tip Warning
If an Error 105 occurs and the instrument is not titled past 7.5 degree a false detection
has occurred. To resolve this issue check the following:
Vibration of the instrument is the most likely cause of a false tip detection. Ensure
the HPQA is on a stable platform
5.
Instrument becomes "stuck" at the lower limit switch
If the instrument stops during a test and becomes "stuck" at the lower limit switch
perform the following
Power cycle the unit by turning power off and on from the switch on the back of the
HPQA
On power-up press the home button
6.
Instrument becomes stuck at the upper limit switch
If the instrument stops at any time and becomes "stuck" at the top limit switch
Turn off power to the unit from the switch on the back of the HPQA
Remove the power cable from the back of the HPQA
Physically push the sled assembly away from the top limit switch. If prior two step are
not completely you will not be able to move the HPQA head assembly
Re-connect the power cable and turn ON the HPQA
On power-up press the home button
7.
Suspected loss of position
If the user suspects that the home position is slipping or has moved from the original
position perform the following
Pres the home button on the LCD and allow the unit to perform a new homing
operation. This will reset the home position
M-HPQA-001
H E L I PAT H Q U I C K A C T I O N L A B O R AT O R Y S TA N D - O P E R AT I O N A L M A N U A L
31