Rear panel

3
Item
1
X1 AC IN connector
2
X2 AC OUT connector
3
X5 HEATER OUT1 connector
4
X7 HEATER OUT2 connector
5
X4 VALVE connector
6
X3 STP connector
7
X6 SENSOR connector
8
Ground terminal (M4)
MT-92E-A01-C
Page 14
1
2
8
1
X1 AC IN connector
2
X2 AC OUT connector
3
X5 HEATER OUT1 connector
4
X7 HEATER OUT2 connector
Figure 12 - STC-U rear panel
A hazardous live voltage may exist at the connectors that are marked with the
warning sign
. DO NOT touch the terminal. Doing so may result in electric
shock. When connecting/disconnecting the connecter, always power off the STC-U
and STP control unit (turn the MAIN POWER "OFF") and isolate (Lockout/Tagout)
the electrical energy source, water and gas, and other energy sources on the
vacuum equipment.
Description
Table 2 - STC-U rear panel connection
STC-U eTMS unit
3
5
5
X4 VALVE connector
6
X3 STP connector
7
X6 SENSOR connector
8
Ground terminal
WARNING
Main power connection (240 Va.c. maximum)
Main power connection to supply to STP control unit (240 Va.c.
maximum)
Main power connection to HEATER 1 (240 Va.c. maximum)
Main power connection to HEATER 2 (240 Va.c. maximum)
VALVE 1 and VALVE 2 connection (24 Vd.c. maximum)
Serial communication to STP control unit and STP pump (12
Vd.c. maximum)
Temperature Sensor connection (5 Vd.c.)