Power Cable; Power Cable (Without Plug) - Edwards STP-XF1303 Series Instruction Manual

Turbomolecular pump
Hide thumbs Also See for STP-XF1303 Series:
Table of Contents

Advertisement

3.6.5

Power cable

Ensure that the STC-U and the power supply cable are suitably protected against earth
(ground) faults and that of "X1 AC IN" connector earth (ground) of the STC-U is
The power cable is designed specifically for the STP pump. DO NOT use the power cable with
Connect the power cable securely to prevent any poor or cross connections. DO NOT apply
Ensure that the supply voltage is as indicated on the STC-U information label.
Connect the power cable to the "X1 AC IN" on the STC-U rear panel as shown in Table 3.
Connect the suitable connector to the power supply as shown in Table 3.
Connect the power cable to the main power of the vacuum equipment via a circuit breaker (rated
current 15 A, with ground fault protection), which complies with UL489 standard and IEC60947-2.
When using the power cable without plug, connect the primary power to the UL-recognized terminal
block of the vacuum equipment. Secure the terminals with M4-bolts as the other side of the terminals,
and cover the terminal block with an appropriate cover. (refer to Figure 21)
19
N
PE( )
L
STC-U side
1
3 pin (socket)
(AWG 14) × 3 cores, φ9.6 to φ10.1 mm
2
2
2.5 mm
Cable color: Brown, Blue, Yellow/Green or Black1, Black2, Yellow/Green
3
Crimp-type terminal (M4)
STC-U eTMS unit
WARNING
correctly connected.
CAUTION
other products.
voltages exceeding 1 kV to the input line.
77
1
Figure 19 - Power cable (without plug)
3
2
Power side (Primary)
MT-92E-A01-C
Page 21