K7GE Trigger Settings
K7GE automatically measures in synchronization with a trigger signal. Set the necessary parameters
for automatic measurement.
This is the Trigger Settings Screen in the
Basic Settings.
Setting parameters are as follows:
Insulation Resistance Monitoring Device
K7GE-MG
For details on setting parameters, hover your mouse over the "i" mark on the screen to display hints
and instructions. Set the parameters according to those instructions.
The K7GE automatic measurement consists of the following four steps. Among these, the time
for Discharge of electric charge and Sampling cannot be changed. For details on operation,
refer to the K7GE-MG User's Manual (N224).
Step: Motor stop standby
Step: Discharge of electric charge
Step: Waiting for stability
Step: Sampling
Condition Monitoring Configuration Tool Usage Guide (N240)
Trigger Signal Reverse, Motor Stop Waiting Time, Time to Wait
to Stabilize, Average Processing
...
Set between 0 and 299 seconds inclusive.
↓
...
Fixed at 20 seconds.
↓
...
Set between 0 and 99 seconds inclusive.
↓
...
Automatic setting
2 Basic Settings
Trigger Settings