102
Error Bitfield byte 1
Error Bitfield byte 2
Access
Type
Description
Read
Unit32
Bit 0 = Leak detected, Leak detect signal high
requires clearing and acknowledging before
pump can resume.
Bit 1 = Motor Stall error active, If set, pump
has a Motor Stall Error. Follow onscreen
instructions
Bit 2 = Motor Speed error. If set pump has a
speed error. Follow onscreen instructions
Bit 3 = Over Current error active. If set, pump
has an over current error. Follow onscreen
instructions
Bit 4 = Over voltage error active. If set, pump
has an over voltage error. Follow onscreen
instructions
Bit 5 = Unused
Bit 6 = Reserved
Bit 7 = Reserved
Unit32
Bit 0 = Reserved
Bit 1 = Reserved
Bit 2 = Reserved
Bit 3 = Reserved
Bit 4 = Reserved
Bit 5 = Reserved
Bit 6 = Reserved
Bit 7 = Reserved