AMI Hydrogen QED
Maintenance
6.5.4
44
Carry out comparison measurement
The comparison measurement is menu driven. Start by selecting
Quality Assurance in menu 3.5 of the monitor AMI Hydrogen QED.
1 Navigate to menu Maintenance /Quality Assurance.
2 Press [Enter].
3 Follow the dialog on the Display.
Quality Assurance
- Carry out preparations
- Install Inspector
- Sample flow to 10 l/h
<Enter> to continue
Quality Assurance
Value H2
0.05 ppb
25.00 C
Value Temp.
Wait 10 Minutes
<Enter> to continue
Quality Assurance
Value H2
0.05 ppb
25.00 C
Value Temp.
Inspector H2
0.06 ppb
Inspector Temp.
<Enter> to continue
Quality Assurance
Value H2
0.05 ppb
25.00 C
Value Temp.
Inspector
0.06 ppm
Inspector Temp.
<Enter> to continue
Quality Assurance
Max. Dev. H2
Max. Dev. Temp.
Dev. H2
Dev. Temp.
QA-Check succesful
4 Carry out pre test preparations
3.5.5
Connect instruments.
Regulate sample flow to 10 l/h us-
ing the appropriate valve.
5 Wait 10 minutes whilst measure-
3.5.5
ment is running.
Press [Enter] to continue.
6 Read the hydrogen value of the
3.5.3
reference instrument and enter un-
der "Inspector." by using the
[
25.0 C
7 Press [Enter] to confirm.
8 Read temperature value of the ref-
3.5.4
erence instrument and enter un-
der "Inspector Temp." by using the
[
25.0 C
9 Press [Enter] to confirm.
10 Press [Enter] to continue.
The results are saved in QA-
3.5.5
History regardless if successful
0.5 %
or not
0.4 °C
0.1 %
0.4 °C
] or [
] keys.
] or [
] keys.
A-96.250.791 / 040219