Carry Out Comparison Measurement - Swann AMI Hydrogen QED Operator's Manual

Table of Contents

Advertisement

AMI Hydrogen QED
Maintenance
6.5.4
44

Carry out comparison measurement

The comparison measurement is menu driven. Start by selecting
Quality Assurance in menu 3.5 of the monitor AMI Hydrogen QED.
1 Navigate to menu Maintenance /Quality Assurance.
2 Press [Enter].
3 Follow the dialog on the Display.
Quality Assurance
- Carry out preparations
- Install Inspector
- Sample flow to 10 l/h
<Enter> to continue
Quality Assurance
Value H2
0.05 ppb
25.00 C
Value Temp.
Wait 10 Minutes
<Enter> to continue
Quality Assurance
Value H2
0.05 ppb
25.00 C
Value Temp.
Inspector H2
0.06 ppb
Inspector Temp.
<Enter> to continue
Quality Assurance
Value H2
0.05 ppb
25.00 C
Value Temp.
Inspector
0.06 ppm
Inspector Temp.
<Enter> to continue
Quality Assurance
Max. Dev. H2
Max. Dev. Temp.
Dev. H2
Dev. Temp.
QA-Check succesful
4 Carry out pre test preparations
3.5.5
Connect instruments.
Regulate sample flow to 10 l/h us-
ing the appropriate valve.
5 Wait 10 minutes whilst measure-
3.5.5
ment is running.
Press [Enter] to continue.
6 Read the hydrogen value of the
3.5.3
reference instrument and enter un-
der "Inspector." by using the
[
25.0 C
7 Press [Enter] to confirm.
8 Read temperature value of the ref-
3.5.4
erence instrument and enter un-
der "Inspector Temp." by using the
[
25.0 C
9 Press [Enter] to confirm.
10 Press [Enter] to continue.
The results are saved in QA-
3.5.5
History regardless if successful
0.5 %
or not
0.4 °C
0.1 %
0.4 °C
] or [
] keys.
] or [
] keys.
A-96.250.791 / 040219