Activate Swan Quality Assurance Procedure - Swann AMI Hydrogen QED Operator's Manual

Table of Contents

Advertisement

AMI Hydrogen QED
Maintenance
Quality Level
0: Off
1: Trend
2: Standard
3: Crucial
4: User
a) sample temperature must have 25°C +/- 5°C.
Procedure
6.5.1
A-96.250.791 / 040219
max. deviation
temperature [°C]
Off
0.5 °C
0.4 °C
0.3 °C
0–2 °C
The standard workflow contains following procedures:

1 Activate SWAN Quality assurance procedure

2 Pre-test
3 Connect instruments
4 Carry out comparison measurement
5 Completion of the measurement
NOTICE: The procedure should only be carried out through
qualified personnel.
Activate SWAN Quality assurance procedure
Enable quality assurance procedure at each instrument to be veri-
fied by selecting the quality level in menu 5.1.4.
The corresponding submenus are then activated.
NOTICE: The activation is necessary the first time only.
max. deviation
a)
result [%]
Off
10%
5%
5%
0–20%
min. inspection
interval
Off
annual
quarterly
monthly
annual, quarterly,
monthly
41