Dräger Evita 2 Instructions For Use Manual page 95

Intensive care ventilator
Table of Contents

Advertisement

Testing Readiness for Operation
-
Leak'test of childrens
hose set
i
l
Switch on machine.
l
Select paed. ventilation, page 22.
l
Set and keep sealed-as
described for adult hose
set, but with rotary knob -VT. set to 0.04 L.
l
Observe display:
P:
mar
plat
PEEP
mean
xxxxxxxx
.
When a peak pressure, p,,,
of 80 mbar has been
reached, the plateau pressure, pp~a, must be not
more than 15 mbar below pm=.
l
If peak pressure remains below 80 mbar. the plateau
pressure, ppw must be not more than 10 mbar
below p,,.
l
Re-connect test lung to Y-piece.