Operation
Switching off alarm limits:
i
Increase upper alarm limit p
until dashes (- -)
appear in the display = switched off.
2
Reduce lower alarm limit I/
until dashes (--I
appear
in the display = switched off.
Recommended
settings:
Lower alarm limit: 20 % lower than VE
Upper alarm limit: 20 % higher than WE
To detect any leakages early, closer alarm limits must be
set.
I
52