Betrieb
Non-operable
settings
Message
cause
Check P(NP/CPAP
CPAP is above piw.
Check puJint.P
Intermittent PEEP is above pw.
check +PEEP
PEEP is above pm~.
Exp. time too short
Display during ILV: expiration time is below 0.5 s.
Insp. hold inactive
Button was pressed for more than 15 s.
IPPV frequency
Slave?
Disolav during ILV: settiw f on Slave machine
Sigh inactive
Check int. PEEP
diff&s-from
setting f on iaster
machine.
Intermittent PEEP is below PEEP.
Volume not constant
pressure limited
Volume not constant
time limited
Tidal volume VT set cannot be supplied,
because pressure limit is in operation.
Tidal volume VT set cannot be supplied,
because inspiration time is too short.
84
Remedy
Reduce CPAP.
Increase pm=
Increase pm=
Reduce IPPV frequency
Release button.
Check f on Slave machine.
Increase intermittent PEEP
Increase insp. flow V~U, or reduce
f, or increase TI:TE.