Operation
Activating
MMV
I
Keep -.
button pressed until the green LED
??I
remains continuously lit.
2
Screen display:
MMV
3
Set IMV frequency.
Spontaneous breathing can be assisted by pressure
assistance, ASB:
4
Set level of pressure assistance.
Set rate of pressure rise for ASB:
5 With J_
rotary knob
left
-A_
stop
Z about 0 s = very steep
right fl
stop
5 about 2 s = gradual
Select measured values for display, page 53.
Other functions
The sensitivity of synchronisation
for pressure assistance
ASB, can be set with the flow trigger,
page 71.
To detect any increase in dead space ventilation during
the spontaneous breathing phase, set tachypnoea
monitoring,
page 68.
To check whether the set tidal volume Vi is being
applied, set volume warning,
page 67.
_-.
Recalling status page for MMi:-
l
Press B
button briefly