Dräger Evita 2 Instructions For Use Manual page 113

Intensive care ventilator
Table of Contents

Advertisement

Technical Data
Additional functions:
lnspiratory relief valve
Safety valve
Opens if compressed air supply
fails (pressure < 1.2 bar],
enables spontaneous breathing
with filtered ambient air.
Opens the breathing system at
100 mbar
Measured
value display
02 concentration
accuracy
5 to 99 vol.%
+ 3 vol.%
Expiratoly minute volume VE
0 to 99 Umin
accuracy
_+ 0% of measured value
to...eo
about 35
Tidal volume Rs
accuracy
0 to 3999 mL
* 8% of measured value
Frequency f
accuracy
to-w
0 to 150lmin
5 llmin
about 35 s
Insp. breathing gas temperature
accuracy
18 to51oc
i 1°C
Airway pressure-data
0 to 99 mbar
During ventilation
peak pressure
max
plateau pressure
plat
pos.endexp.pressure
PEEP
mean pressure
mean
During spontaneous breathing
max. airway pressure
max
continuous positive
airway pressure
CPAP
min. airway pressure
min
During BIPAP
max. airway pressure
maz
continuous positive
airway pressure
CPAP
min. airway pressure
min
mean pressure
mean
accuracy
* 2 mbar