Fitting 02 Sensor - Dräger Evita 2 Instructions For Use Manual

Intensive care ventilator
Table of Contents

Advertisement

Preparation
Fitting 02 sensor
-
when operating for the first time
-
when display reads:
01 meas.
hop
when calibration can no longer be carried out.
1
Turn port downwards
or to the left.
2
Use coin to loosen screw, remove protective cover.
3
Remove 02 sensor from the mounting and unscrew
housing.
Insert new sensor
capsule:
l
The side with circular contacts goes into the housing
l
Re-assemble sensor housing, push firmly into the
mounting. Re-assemble protective cover.
l
Dispose of used sensor, page 102.