Dräger Evita 2 Instructions For Use Manual page 89

Intensive care ventilator
Table of Contents

Advertisement

Care
Wipe disinfection
Basic machine without hose system,
Aquapor without bowl and float,
gas-connecting
hoses,
temperature sensor
l
Disinfect by wiping with Buraton (Messrs. Schijlke &
Mayr, Norderstedt),
for instance.
Observe manufacturer's
instructions.
or:
Disinfect
in Aseptor
disinfection
chamber
Basic machine without hose system, without cooling-air
filter
and without ambient-air filter,
Aquapor without bowl and float,
'gasconnecting
hoses,
temperature sensor.
Remove obvious stains first with disposable cloth.
Remove cooling-air filter, page 102.
Remove ambient-air filter, page 103.
Dry parts well; damp parts cause unpleasant smells.
Place basic machine, Aquapor and accessories in
Aseptor and disinfect in accordance with Aseptor
Instructions for Use.
Afterwards:
l
Replace cooling-air filter, page 10'2
and
l
Replace ambient-air filter, page 103.
Disinfection
with moist heat
All disassembled parts:
Ventilation hoses,
water traps and their collecting jars,
Aquapor bowl with float and socket,
Y-piece, medicament nebulizer, patient system or their
parts, if these are badly soiled.
.
.
Disinfect with moist heat (93'C/lO
minutes) in a
cleaning and disinfecting machine.
Use only cleaning
agent.
After disinfection with moist heat steam sterilize
patient system and individual
parts at 134°C to
make sure that no water remains in the pressure
measuring line and possibly causes some malfunction.