Maintenance Intervals; Disposal Of Nicd Batteries - Dräger Babytherm 8000 OC Instructions For Use Manual

Resuscitation/open care unit
Table of Contents

Advertisement

Maintenance

Disposal of NiCd batteries

Maintenance Intervals

The unit or its components have to be disinfected and
cleaned prior to each maintenance procedure, and
also before unit is returned for repair.
Disconnect from mains power supply before each
[
maintenance procedure.
Gel mattress
Cover small cracks in the foil
with adhesive tape from the
repair set.
Flexible hose seals
Replace when brittle and sticky.
Temperatur measuring Check the measurements at
device
the same time as other technical
(radiant heater)
safety checks are carried out by
trained personnel.
NiCd battery for mains to be replaced by trained service
failure alarm
personnel after 2 years.
(radiant heater)
Time Keeper RAM
to be replaced by trained service
personnel after 4 years.
Inspection and
yearly by trained service
maintenance*
personnel.
Disposal of NiCd batteries
– do not throw into fire, risk of explosion.
– do not force open, danger from corrosive substances.
Batteries must be disposed of as special waste:
Follow local regulations on the disposal of waste
products.
For further information contact the local authority or
Environmental Control Office, or approved waste
disposal contractors.
* Definitions:
Inspection
= examination of actual condition
Service
= measures to maintain specified condition
Repair
= measures to restore specified condition
Maintenance
= inspection, service, repair