Specifications - Masterflex MFLX77112-10 Operating Manual

Ss b/t washdown digital pump drives with advanced connectivity
Table of Contents

Advertisement

Section 5: Service & Maintenance

SPECIFICATIONS

OUTPUT
Speed:
Torque Output,
Max Continuous:
Speed Regulation:
Display:
Outputs:
INPUT
Supply Voltage Limits:
Max Current:
Inputs:
CONSTRUCTION
Dimensions (L x W x
H):
Weight:
5-8
Masterflex SS B/T Washdown Digital Pump Drive with Advanced Connectivity
11 to 321 RPM
.4 HP x 63025 /321 RPM= 78.5 in-lbs
78.5 in-lbs x16= 1256.6 oz-in
Line ±0.1% F.S.
Load ±0.1% F.S.
Drift ±0.1% F.S.
800 x 480 LCD w/ LED Backlight
Voltage speed output (0-10 VDC @ 1 k min)
Current speed output
(4-20 mA @ 0-600 )
Motor running relay
(Closed when running, 1 A max @ 24 VDC)
General alarm relay
(Closed when error is displayed, 1 A @ 24 VDC)
Open head relay
(Closed when activated, 1 A @ 24 VDC)
Tach output (110 – 3210 Hz, 50% Duty Cycle, 10 Hz
per RPM)
90 to 230 Vrms @ 50/60 Hz (Universal Input)
5.0 A @ 115 VAC; 3.1 A @ 230 VAC
Start/Stop, Prime, CW/CCW
(Contact closure)
Current input
(0-20 mA @ 249 ; ± 50 V common mode range)
Voltage input
(0-10 VDC @ 10 k; ± 50 V common mode range)
RJ45 Ethernet
USB, Type A
(5 VDC @ 0.5 A)
24.5" x 15.8" x 18.8"
(622 mm x 401 mm x 478 mm)
120 lbs. (54.4 kg)
Masterflex