Page 2
US and in other countries. MasterSense™ EtherNet/IP™ is a trademark of ODVA Windows is a registered trademark of Microsoft PUMP FOR LIQUIDS ORIGINAL INSTRUCTIONS ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Saving Continuous Mode Settings as a New Program Time Mode Time Mode Run Screen Time Mode Edit Screen Time Mode Operation Saving Time Mode Settings as a New Program Volume Mode ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 4
SECTION 5: SERVICE & MAINTENANCE Firmware Updates Restore Factory Settings Cleaning the Pump Drive Accessories Replacement Parts Troubleshooting Error Defi nitions DB-25 Electrical Connections Specifi cations Technical Assistance Product Return Warranty Disposal ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 5
® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
CAUTION: It is recommended that any repairs be performed only by an authorized technician. If service and repairs are performed by the customer or by any third party company, Masterfl ex denies all responsibility. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
L/S models have a maximum speed of 600 RPM and fl ow rate capacities that range from .001 mL/min through to 3400 mL/min (depending on the model and tubing size selected). ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Section 2: Basic Setup & Settings SECTION 2: BASIC SETUP & SETTINGS Touchscreen Active Light Pump Head Mounting Holes Power Switch Ethernet Port USB Port DB-25 Open Head Sensor Profibus Power Cord Inlet ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Confirm Ramp Up Connectivity Status Record Volume Continuous Mode Reset Counterclockwise Screen Lock Delete Settings Display Brightness Start Edit Stop Language Time Mode Logout Update Available New Program Volume Mode ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
After 30 minutes of inactivity the L/S will enter sleep mode and the display will turn off . Tapping the touchscreen will reactivate the drive. To enable (default) or disable the screen saver see "Screen Saver" on page 2-6. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
5. Tap MM to select minutes and enter the desired time using the onscreen keypad. 6. If using 12-HOUR time, tap AM/PM to select either AM or PM. 7. Tap CONFIRM to save or CANCEL to discard changes. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Select the desired current or voltage operating range from the available list. c. Tap CONFIRM to save or CANCEL to discard changes. Th e Analog Output Screen will be displayed. 5. Tap CONFIRM ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
3. Tap DEVICE NAME. Th e Device Name Screen will be displayed. 4. Enter the desired pump name using the onscreen keypad (up to 8 characters). 5. Tap CONFIRM to save or CANCEL to discard changes. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
3. Tap ETHERNET SETTINGS. Th e Ethernet Settings Screen will be displayed. 4. Tap DHCP to automatically confi gure IP settings or STATIC to manually edit settings using the onscreen keypad. 5. Tap CONFIRM to save or CANCEL to discard changes. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
8. Tap the required PERMISSION LEVEL. NOTE: Th ere must always be at least one ADMIN user. 9. Tap CONFIRM to save or CANCEL to discard any changes. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
2. Select the desired tube size from the list of available options. NOTE: Only tube sizes compatible with the selected pump head will be displayed. 3. Tap CONFIRM to save or CANCEL to discard changes and return to the mode run screen. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
5. Navigate to the Continuous, Time or Volume Mode screens. 6. Press and hold PRIME until there are no bubbles visible in the tubing. Priming will stop when PRIME is released. ® Masterflex 2-10 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
(For further information see "User Management" on page 2-8). 4. Navigate to the Continuous, Time or Volume Mode screens. 5. Tap DIRECTION to select either clockwise or counterclockwise fl ow direction. ® Masterflex 2-11 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 19
Th e best results are obtained after tubing has been run in the pump for approximately 20-30 minutes. • Steps 9–14 can be repeated as necessary to optimize the accuracy of the tubing calibration. ® Masterflex 2-12 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Return to Previous Screen Continuous Mode Time Mode Volume Mode Analog Input Mode Page Scroll Add New Program Settings Tap the desired mode or program to access individual mode or program screens. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
UNLOCKED until the icon changes to LOCKED and a red border appears around the screen. To unlock the screen: Press and hold LOCKED until the icon changes to UNLOCKED ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
8. Tap STOP when the drive operation is no longer required. NOTE: In Continuous Mode the drive will continue to operate at the displayed fl ow rate and direction until stopped. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
• Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. • If there are multiple programs tapping SCROLL will display additional pages. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
J. PRIME: Press and hold to prime the pump drive (for further information see "Priming the Pump Drive" on page 2-10). K. START/STOP/PAUSE: During operation, the display will change from START to PAUSE STOP/ RESET ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Time Mode Edit Screen (for further information see "Tube Calibration" on page 2-11). • If required, prime the pump before operation (for further information see "Priming the Pump Drive" on page 2-10). ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 26
11. If required, adjustments can be made to any of the Time Mode option settings during operation. To make any adjustments: a. Tap PAUSE and then repeat steps 2–9 above. b. Tap CONTINUE to complete the pump operation once the desired changes have been made. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
• Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. • If there are multiple programs tapping SCROLL will display additional pages. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
I. FLOW RATE: Displays the current fl ow rate in the unit of measurement selected by the user. J. PRIME: Press and hold to prime the pump drive (for further information see "Priming the Pump Drive" on page 2-10). ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Confi rm that the tubing has been calibrated by checking that Calibration Complete is displayed on the Volume Mode Edit Screen (for further information see "Tube Calibration" on page 2-11). ® Masterflex 3-10 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 30
Volume Mode Edit Screen. 8. If required, tap ANTI-DRIP. Th e Anti-Drip Screen will be displayed (for further information see "Anti- Drip" on page 3-12). ® Masterflex 3-11 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Tap FLOW to enter the desired fl ow rate using the onscreen keypad. b. Tap UNITS to enter the desired fl ow rate units using the onscreen keypad. c. Tap CONFIRM to save changes and return to the Record Dispense Screen. ® Masterflex 3-12 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Th e Analog Input Mode Run Screen is accessed by selecting ANALOG INPUT from the Mode Selection Screen. A. Mode Name Banner B. Prime C. Start/Stop D. Flow Direction E. Screen Lock F. Logout ® Masterflex 3-13 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
If required, prime the pump before operation (for further information see "Priming the Pump Drive" on page 2-10). • Confi rm that the pump drive is connected to the appropriate equipment through the drive’s DB-25 female connection port. ® Masterflex 3-14 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. • If there are multiple programs tapping SCROLL will display additional pages. ® Masterflex 3-15 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. • If there are multiple programs tapping SCROLL will display additional pages. Adding a New Program: Time Mode ® Masterflex 3-16 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 36
10. Tap CONFIRM to save or DELETE to discard changes. NOTE: • Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. ® Masterflex 3-17 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
8. Tap DIRECTION to select either clockwise or counterclockwise fl ow direction. 9. Tap BATCH TOTAL to adjust the number of dispenses in each batch cycle. Th e Batch Total Screen will be displayed. ® Masterflex 3-18 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Analog Input New Program Screen. 8. Tap DIRECTION to select either clockwise or counterclockwise fl ow direction. 9. Tap CONFIRM to save or DELETE to discard changes. ® Masterflex 3-19 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
2. Edit the settings for the selected mode as desired. For instructions on editing individual mode settings see: • "Adding a New Program: Continuous Mode" on page 3-16. • "Adding a New Program: Time Mode" on page 3-16. ® Masterflex 3-20 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Th e Program Edit Screen will be displayed. 2. Tap DELETE PROGRAM 3. Tap DELETE to delete the program and return to the Mode Selection Screen or CANCEL to return to the previous screen. ® Masterflex 3-21 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
IP address is within the subnet mask range of the PLC and the pump. Th e default addressing methods for the pump is DHCP. Static IP address may be assigned through the Ethernet settings menu. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 42
Section 4: Communication Specification In the device menu, enable the Ethernet/IP feature by toggling to select ON. Once enable, the following message appears on the UI. Press Accept on this screen. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 43
-Th e bit changes state from local to remote on a 1 to 0 transition. -Once enabled, the Ethernet/IP banner will appear at the top of the pump screen. Shows Ethernet/IP active in volume Dispense Mode ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Section 4: Communication Specification MASTERFLEX ETHERNET/IP Input data; 56 bytes of input data from pump to master. Bytes Data Type Description 32-Bit INT Pump Status Bit 0: Status OK Bit 1: Pump Running Bit 2: Dispense Running Bit 3: Tube Uncalibrated...
Page 45
Set Batch Count Total (0 = infinite) Masterfl ex Flow Units Correlation Table Index 600 & 100 RPM L/S mL/min mL/hr L/min L/hr L/day uL/min uL/hr gal/min gal/hr gal/day oz/min oz/hr cum/hr ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Once that fi eld is selected you now enter the PROFIBUS SETTINGS screen. Here you can assign the Profi bus address to be a value (in between 1 and 126). Th e default Profi bus address is 7. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 47
Ensure the pump and the PLC user interface are confi gured for the same address and confi rm by pressing the green check mark at the top of the screen ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Section 4: Communication Specification PROFIBUS DATA Th e MASTERFLEX Pump Drive operates as a DP-V0 Slave. GSD File: CPMP0FF8.GSD ID Number: 0FF8 HEX Default Slave Address: 7 Cyclic Input Data: 56 bytes Cyclic Output Data: 28 bytes Table 1 below lists the 56 bytes of cyclic input data, from pump to master.
Page 49
Pump Control, a 4-byte unsigned integer as shown in Table 3 above, controls the pump’s operation in separate bits. See Table 4 below for the description of each bit in Pump Control. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 50
Set Dispense On Seconds is the amount of time the pump will dispense in Time Dispense Mode. Set Dispense Off Seconds is used for both Time and Volume dispense modes. It sets the amount of time the pump will pause between dispenses. ® Masterflex 4-10 MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Th e Settings Screen will be displayed. 2. SCROLL through the Settings Screen pages to locate DEVICE INFORMATION. 3. Tap DEVICE INFORMATION. 4. Tap FACTORY RESET and follow the onscreen prompts. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Wearing thick latex Remove gloves before using the touchscreen. NOTE: Consult sive. gloves can make the your company or laboratory safety guidelines before removing screen less responsive. any personal protective equipment. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
The drive has exceeded commanded speed value. Actions: Drive will stop immediately. Verify load is correct and power cycle the pump drive. If error persists see "Technical Assistance" on page 5-8. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
(5 VDC @ 0.5 A) CONSTRUCTION Dimensions (L x W x H): 10.12” x 7.62” x 8.51” (258 mm x 194 mm x 216 mm) Weight: 12.5 lbs. (5.66 kg) ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 57
Exposed material is stainless steel, aluminum, thermoplastic COMPLIANCE All Models UL 61010-1, UL 61010-2-081 US/CAN For CE mark: EN61010-1, Low Voltage Directive EN61326-1, EMC Directive EN50581, RoHS Directive Pump Head EN809, Machinery Directive ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Once the useful life of the product has ended, please ensure proper disposal according to local laws. Plastic and electronic components should be disposed of at a recycling facility. Please refer to local regulations regarding proper disposal. ® Masterflex MasterSense™ L/S Digital Pump Drive with Advanced Connectivity...
Page 59
US & Canada only *EN809 manufactured by: Toll Free 1-800-MASTERFLEX | 1-800-637-3739 Masterflex LLC 28092 W Commercial Avenue, Barrington, IL 60010 Outside US & Canada masterflex.tech @ avantorsciences.com 1-847-381-7050 www.avantorsciences.com/masterflex...