Profibus Data - Masterflex MFLX77112-10 Operating Manual

Ss b/t washdown digital pump drives with advanced connectivity
Table of Contents

Advertisement

Section 4: Communication Specification

PROFIBUS DATA

Th e MASTERFLEX Pump Drive operates as a DP-V0 Slave.
GSD File: CPMP0FF8.GSD
ID Number: 0FF8 HEX
Default Slave Address: 7
Cyclic Input Data: 56 bytes
Cyclic Output Data: 28 bytes
Table 1 below lists the 56 bytes of cyclic input data, from pump to master.
Table 1
Cyclic Input (pump —> master, 56 bytes)
Data
Pump Status
Dispense Mode
Tube Size
Flow Units
Padding (not used)
Cumulative Volume
Remaining Dispense Volume
Remaining Dispense On Seconds
Remaining Dispense Off Seconds
Current Batch Count
Total Batch Count
Minimum Flow Rate
Current Flow Rate
Maximum Flow Rate
Remaining Dispense On Days
Remaining Dispense On Hours
Remaining Dispense On Minutes
Remaining Dispense On Seconds
Remaining Dispense On Tenths
Remaining Dispense Off Days
Remaining Dispense Off Hours
Remaining Dispense Off Minutes
Remaining Dispense Off Seconds
Remaining Dispense Off Tenths
4-8
Masterflex SS B/T Washdown Digital Pump Drive with Advanced Connectivity
Size in Bytes Data Type
4
unsigned integer
1
unsigned integer
1
unsigned integer
1
unsigned integer
1
unsigned integer
4
decimal floating-point
4
decimal floating-point
4
decimal floating-point
4
decimal floating-point
4
unsigned integer
4
unsigned integer
4
decimal floating-point
4
decimal floating-point
4
decimal floating-point
2
unsigned integer
1
unsigned integer
1
unsigned integer
1
unsigned integer
1
unsigned integer
2
unsigned integer
1
unsigned integer
1
unsigned integer
1
unsigned integer
1
unsigned integer
Masterflex

Advertisement

Table of Contents
loading

Table of Contents