Masterflex ISMATEC 78018-50 Operating Manual

Reglo digital pump drives with advanced connectivity
Table of Contents

Advertisement

MASTERFLEX
REGLO 78018-60
®
MASTERFLEX
REGLO 78018-50
®
(US & Canada only) Toll Free 1-800-MASTERFLEX • 1-800-637-3739
(Outside US & Canada) 1-847-549-7600 • 1-847-381-7050
www.masterflex.com • techinfo@masterflex.com
OPERATING MANUAL:
REGLO DIGITAL PUMP
DRIVES WITH ADVANCED
CONNECTIVITY
Model Nos.
78018-50
78018-60
A-1299-5202
Edition 01

Advertisement

Table of Contents
loading

Summary of Contents for Masterflex ISMATEC 78018-50

  • Page 1 REGLO DIGITAL PUMP DRIVES WITH ADVANCED CONNECTIVITY Model Nos. 78018-50 78018-60 MASTERFLEX REGLO 78018-60 ® MASTERFLEX REGLO 78018-50 ® A-1299-5202 Edition 01 (US & Canada only) Toll Free 1-800-MASTERFLEX • 1-800-637-3739 (Outside US & Canada) 1-847-549-7600 • 1-847-381-7050 www.masterflex.com • techinfo@masterflex.com...
  • Page 2 © 2019 Cole-Parmer Instrument Company. All rights reserved. Masterflex® is a registered trademark of the Cole-Parmer Instrument Company. Viton® is a registered trademark of the Chemours Company. Trademarks bearing the symbol in this publication are registered in the US and in other countries.
  • Page 3: Table Of Contents

    Priming the Gear Pump 3-11 Priming the Piston Pump 3-11 Pump Head Calibration 3-12 Pump Head Calibration Screen 3-12 Gear Pump Calibration 3-12 Piston Pump Calibration 3-14 SECTION 3: OPERATION 3-16 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 4 Flushing the Pump Head 3-43 Cleaning the Pump Drive 3-43 Accessories 3-43 DB-25 Electrical Connections 3-44 Pump Head Replacement 3-45 Available A-Mount Gear Pump Heads 3-45 Available Piston Pump Heads 3-46 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 5 Table of Contents Mounting a Ceramic Piston Pump 3-47 Mounting a Gear Pump 3-49 Troubleshooting 3-51 Error Definitions 3-52 Specifications 3-53 Technical Assistance 3-55 Product Return 3-55 Warranty 3-55 Disposal 3-55 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 6: Section 1: Introduction

    CAUTION: Do not handle the pump head before the pump is switched off and disconnected from the mains power supply. CAUTION: The pump must be switched off before removing the pump head. Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 7: Media Containing Particles

    While priming, ensure the pump does not run dry for more than 15 seconds. Pump Against Pressure The REGLO piston pump system can be used for continuous operation at a maximum differential pressure of 6.9 bar. CAUTION: Please consult factory for maximum pressure capabilities. Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 8: About The Reglo

    The REGLO range features Ethernet and Wi-Fi connections for web-based control and monitoring using the MasterflexLive™ website (see www.masterflex.com), and remote analog input control and monitoring through the pump drive’s DB-25 female connection port.
  • Page 9: Peristaltic Cassette Models

    Click‘n’go 78018-14 2CH12RLR Peristaltic Click’n’go 78018-20 4CH6RLR Peristaltic Click’n’go 78018-22 4CH8RLR Peristaltic Click’n’go 78018-24 4CH12RLR Peristaltic Click’n’go 78018-40 Miniflex Peristaltic Miniflex Pump (Single Channel) 78018-42 Miniflex Peristaltic Miniflex Pump (Dual Channel) Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 10: Section 2: Basic Setup & Settings

    Flow Ring Fixing Screw Pump Head Piston Pump Model 78018-60 Touchscreen Fixing Case Pump Head Tubing Adapters Gear Pump Model 78018-50 DB-25 Port Power Switch Power Cable Locking Barrel Connector Ethernet Port USB Port Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 11: Touchscreen Icons

    Ramp Down Confirm Ramp Up Connectivity Status Record Volume Continuous Mode Reset Counterclockwise Screen Lock Delete Settings Display Brightness Start Edit Stop Language Time Mode Logout Update Available New Program Volume Mode Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 12: Before Using The Drive

    Recalibrate the pump head regularly to maintain the best flow rate accuracy (for further information see "Pump Head Calibration" on page 3-12). • For tubing selection and compatibility go to www.masterflex.com. • Remove the pump drive’s power cord from the mains power outlet when cleaning or performing maintenance on the drive.
  • Page 13: Pump Preparation

    After 30 minutes of inactivity the REGLO will enter sleep mode and the display will turn off. Tapping the touchscreen will reactivate the drive. To enable (default) or disable the screen saver see "Auto Start" on page 3-10. Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 14: Settings

    5. Tap MM to select minutes and enter the desired time using the onscreen keypad. 6. If using 12-HOUR time, tap AM/PM to select either AM or PM. 7. Tap CONFIRM to save or CANCEL to discard changes. Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 15: Setting The Date

    Select the desired current or voltage operating range from the available list. c. Tap CONFIRM to save or CANCEL to discard changes. The Analog Output Screen will be displayed. 5. Tap CONFIRM Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 16: Device Information

    3. Tap DEVICE NAME. The Device Name Screen will be displayed. 4. Enter the desired pump name using the onscreen keypad (up to 8 characters). 5. Tap CONFIRM to save or CANCEL to discard changes. Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 17: Wi-Fi Settings

    3. Tap ETHERNET SETTINGS. The Ethernet Settings Screen will be displayed. 4. Tap DHCP to automatically configure IP settings or STATIC to manually edit settings using the onscreen keypad. 5. Tap CONFIRM to save or CANCEL to discard changes. Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 18: User Management

    4. Locate the desired username from the available list and then tap the DELETE USER icon located next to the username. 5. Tap DELETE to delete user or CANCEL to cancel. Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 19: Auto Start

    The Settings Screen will be displayed. 2. SCROLL through the Settings Screen pages to locate AUTO START. 3. Tap the AUTO START toggle to select either ON or OFF. 3-10 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 20: Priming The Pump

    6. Navigate to the Continuous, Time or Volume Mode screens. 7. Press and hold PRIME until no bubbles are visible in the tubing. Priming will stop when PRIME is released. 3-11 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 21: Pump Head Calibration

    3. Switch the drive on using the power switch located at the rear of the drive. NOTE: If Manage User is enabled you may be prompted for a Username and Password (for further information see "User Management" on page 3-9). 3-12 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 22 Minimum and maximum flow rates will change after calibration due to a recalculation of the volume of fluid per pump revolution. • Steps 6–12 can be repeated as necessary to optimize the accuracy of the pump head calibration. 3-13 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 23: Piston Pump Calibration

    Calibration Complete should now be displayed to indicate that calibration has been completed for the current pump head size, tube volume, and flow rate. 3-14 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 24 Minimum and maximum flow rates will change after calibration due to a recalculation of the volume of fluid per pump revolution. • Steps 8–14 can be repeated as necessary to optimize the accuracy of the pump head calibration. 3-15 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 25: Section 3: Operation

    Before operating the pump it is recommended that the pump head be filled with liquid. • While priming, ensure the pump does not run dry for more than 15 seconds. 3-16 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 26: Mode Selection Screen

    Return to Previous Screen Continuous Mode Time Mode Volume Mode Analog Input Mode Page Scroll Add New Program Settings Tap the desired mode or program to access individual mode or program screens. 3-17 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 27: Continuous Mode

    G. START/STOP: Tap START to begin operation. Once operation has commenced STOP will be displayed. H. FLOW DIRECTION: Tap DIRECTION to select either clockwise or counterclockwise NOTE: Flow Direction is unavailable on single direction gear pumps and will be grayed out. 3-18 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 28: Continuous Mode Operation

    8. Tap STOP when the drive operation is no longer required. NOTE: In Continuous Mode the drive will continue to operate at the displayed flow rate and direction until stopped. 3-19 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 29: Saving Continuous Mode Settings As A New Program

    • Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. • If there are multiple programs tapping SCROLL will display additional pages. 3-20 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 30: Time Mode

    I. FLOW RATE: Displays the current flow rate in the unit of measurement selected by the user. J. PRIME: Press and hold to prime the pump drive (for further information see "Priming the Pump" on page 3-11). 3-21 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 31: Time Mode Edit Screen

    Time Mode Edit Screen (for further information see "Pump Head Calibration" on page 3-12). • If required, prime the pump before operation (for further information see "Priming the Pump" on page 3-11). 3-22 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 32 11. If required, adjustments can be made to any of the Time Mode settings during operation. To make any adjustments: a. Tap PAUSE and then repeat steps 2–9 above. b. Tap CONTINUE to complete the pump operation once the desired changes have been made. 3-23 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 33: Saving Time Mode Settings As A New Program

    • Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. • If there are multiple programs tapping SCROLL will display additional pages. 3-24 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 34: Volume Mode

    I. FLOW RATE: Displays the current flow rate in the unit of measurement selected by the user. J. PRIME: Press and hold to prime the pump drive (for further information see "Priming the Pump" on page 3-11). 3-25 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 35: Volume Mode Edit Screen

    Volume Mode Edit Screen (for further information see "Pump Head Calibration" on page 3-12). • If required, prime the pump before operation (for further information see "Priming the Pump" on page 3-11). 3-26 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 36 8. If required, tap ANTI-DRIP. The Anti-Drip Screen will be displayed (for further information see "Anti- Drip" on page 3-29). NOTE: Anti-Drip is unavailable on gear pumps. a. Tap ON or OFF to enable or disable the Anti-Drip feature. 3-27 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 37: Saving Volume Mode Settings As A New Program

    • Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. • If there are multiple programs tapping SCROLL will display additional pages. 3-28 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 38: Anti-Drip

    NOTE: If required, tap STOP and then CANCEL to discard recording and return to the Volume Mode Edit Screen. 3-29 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 39 Volume Mode Edit Screen. If saved, the recorded dispense volume will now be displayed under VOLUME. 9. Tap CONFIRM . The Volume Mode Run Screen will be displayed. 3-30 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 40: Analog Input Mode

    UNLOCKED until the icon changes to LOCKED and a red border appears around the screen. To unlock the screen: Press and hold LOCKED until the icon changes to UNLOCKED 3-31 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 41: Analog Input

    NOTE: If locked, the screen will need to be unlocked before the pump operation can be stopped. • www.masterflex.com for further information on using analog input remote control and monitoring. 3-32 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 42: Saving Analog Input Mode Settings As A New Program

    • Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. • If there are multiple programs tapping SCROLL will display additional pages. 3-33 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 43: Masterflexlive

    Confirm the correct transfer tubing is securely attached to the pump head. • Confirm all fluids and containers are ready. • Switch the drive on using the power switch located at the rear of the drive. NOTE: If Manage User 3-34 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 44 • Tap STOP on the pump drive touchscreen to stop the pump operation. NOTE: If locked, the screen will need to be unlocked before the pump operation can be stopped. 3-35 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 45: Custom Program Modes

    2. Tap TIME. The Time New Program Screen will be displayed. 3. Tap PROGRAM1. If required, delete the characters using the back arrow on the onscreen keypad and then enter the desired program name (up to 8 characters). 3-36 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 46: Adding A New Program: Volume Mode

    1. Tap NEW PROGRAM from the Mode Selection Screen. The New Program Screen will be displayed. NOTE: If NEW PROGRAM is not displayed on the Mode Selection Screen tap SCROLL to view additional pages. 2. Tap VOLUME. The Volume New Program Screen will be displayed. 3-37 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 47 Tap ON or OFF to enable or disable the Anti-Drip feature. b. If enabled, enter the desired degrees of reverse rotation using the onscreen keypad. NOTE: Typical values range from 5 to 45 degrees. 3-38 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 48: Adding A New Program: Analog Input Mode

    • Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection Screen. • If there are multiple programs tapping SCROLL will display additional pages. 3-39 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 49: Using Program Modes

    NOTE: Once the individual mode settings have been edited and saved the Program Edit Screen will be displayed. 3. Tap CONFIRM to save or CANCEL to discard changes and return to the Mode Selection Screen. 3-40 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 50: Deleting A Program

    The Program Edit Screen will be displayed. 2. Tap DELETE PROGRAM 3. Tap DELETE to delete the program and return to the Mode Selection Screen or CANCEL to return to the previous screen. 3-41 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 51: Section 4: Service & Maintenance

    The Settings Screen will be displayed. 2. SCROLL through the Settings Screen pages to locate DEVICE INFORMATION. 3. Tap DEVICE INFORMATION. 4. Tap FACTORY RESET and follow the onscreen prompts. 3-42 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 52: Flushing The Pump Head

    Do not operate the pump drive with water on the surface of the touchscreen. • Any damage caused by improper cleaning will be the sole responsibility of the customer. ACCESSORIES To purchase accessories or for further information visit www.masterflex.com. The following accessories are available for the REGLO Digital Pump Drive: Description Part Number Footswitch with 6 ft (1.8 m) cable and DB-25 male connector...
  • Page 53: Db-25 Electrical Connections

    Start/Stop Tach Output CW/CCW Prime Output (0–20 mA, 4–20 mA) Motor Running (N.O.) Input (0–20 mA, 4–20 mA) Open Head Sensor (N.O.) Input (0–10 V) General Alarm Output (0–10 V) 3-44 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 54: Pump Head Replacement

    73011-02 0.32 316 SS PEEK PTFE with internal bypass 73011-08 0.64 316 SS PEEK PTFE 73011-16 0.91 316 SS PEEK PTFE 73011-18 0.91 316 SS PEEK PTFE with internal bypass 3-45 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 55: Available Piston Pump Heads

    1/4 -28 Female 37.5 Ceramic, ETFE 07117-63 1/4" OD Tubing Ceramic, ETFE 07117-65 1/4" OD Tubing Ceramic, PVDF 07117-69 1/4 -28 Female Ceramic, PVDF 07117-71 1/4 -28 Female Ceramic, ETFE 3-46 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 56: Mounting A Ceramic Piston Pump

    Place the coupling onto the pump-head shaft and push it on as far as it will go. NOTE: The Allen screw must be centered on the flat of the pump-head shaft. b. Tighten the coupling (2) with the 2.5 mm size Allen key (D). 3-47 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 57 "Priming the Piston Pump" on page 3-11). 9. Calibrate the new pump head (for further information see "Piston Pump Calibration" on page 3-14). The new pump head is now ready for operation. 3-48 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 58: Mounting A Gear Pump

    Coupling • During operation the back of the coupling magnet must Magnet not touch the housing of the pump drive. Allen Key 2.5 mm (3/32") 3-49 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 59 "Gear Pump Calibration" on page 3-12). The new pump head is now ready for operation. CAUTION: Fill the pump head with liquid before pump operation. Avoid dry running the pump head. 3-50 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 60: Troubleshooting

    Wearing thick latex Remove gloves before using the touchscreen. NOTE: Consult gloves can make the your company or laboratory safety guidelines before removing any screen less responsive. personal protective equipment. 3-51 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 61: Error Definitions

    Bad EEPROM checksum for Factory Cal. Actions: Error will be cleared after 10 seconds and parameters will be reset to default values. If error persists see "Technical Assistance" on page 3-55. 3-52 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 62: Specifications

    USB, Type A (5 VDC @ 0.5 A) CONSTRUCTION Dimensions (L x W x H): Drive Only 7.62” x 5.75” x 6.31” (194 mm x 146 mm x 160 mm) 3-53 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 63 Exposed material is painted aluminum and ABS plastic COMPLIANCE All Models UL 61010-1 US/CAN For CE mark: EN61010-1, Low Voltage Directive EN61326-1, EMC Directive EN50581, RoHS Directive Pump Head Models EN809, Machinery Directive 3-54 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 64: Technical Assistance

    Once the useful life of the product has ended, please ensure proper disposal according to local laws. Plastic and electronic components should be disposed of at a recycling facility. Please refer to local regulations regarding proper disposal. 3-55 Masterflex REGLO Digital Pump Drive with Advanced Connectivity Masterflex...
  • Page 65 US & Canada only *EN809 manufactured by: Toll Free 1-800-MASTERFLEX | 1-800-637-3739 Cole-Parmer Instrument Company 28W092 Commercial Avenue, Barrington, IL 60010 Outside US & Canada techinfo@masterflex.com | www.masterflex.com 1-847-549-7600 | 1-847-381-7050...

This manual is also suitable for:

Ismatec 78018-60

Table of Contents