Display
An audible signal sounds.
The three green status LEDs of
the function check light up.
Two signal tones sound.
The red status LED lights up
and one or more green status
LEDs of the failed tests flash.
The green status LEDs of the
passed tests light up.
Result
The automatic function check starts with the audible signal and all
the LEDs on the control panel light up briefly. The LEDs of the On/
Off button
flash during the automatic function check. The
automatic function check includes the following sequence of tests:
Leaktightness
System
Battery
6. To cancel the function check: Press the test button
or
Press the On/Off button
7. Proceed with the device according to the following table:
Meaning
Function check passed
Function check failed
8. If necessary: Switch off the device
device",
page 62).
The automatic function check is complete.
.
Action
Use device without restriction.
Take action
function check (ACCUVAC Pro)",
page 92).
(see „4.7 Switching off the
ACCUVAC Pro/ACCUVAC Lite
6 Function check
(see „6.2.3 Failed
EN
91