Weinmann ACCUVAC Pro Instructions For Use Manual page 106

Suction device
Hide thumbs Also See for ACCUVAC Pro:
Table of Contents

Advertisement

11 Appendix
Specification
Article number (REF)
Charging process:
Temperature range
Humidity
Air pressure
Repeated technical safety
checks (STKs, only in Germany)
11.1.2 Technical data for battery
Specification
Type
Dimensions (W x H x D)
Weight
Nominal capacity
Rated voltage
Charging time
Charging interval in long-term storage
Battery run time in continuous
operation with fully charged/new battery
(> 20 l/min, setting -0.8 bar)
Service life
Display
Typical battery life*
106
EN
ACCUVAC Pro/ACCUVAC Lite
ACCUVAC Pro
WM 11602
0 °C to +40 °C
5 % to 95 % relative humidity
without condensation
540 hPa to 1100 hPa
Not applicable
*1 bar = 100 kPa
Subject to alterations in design
ACCUVAC Pro
Li-ion
4.3 x 7.3 x 7.5 cm
0.4 kg
Min. 4.3 Ah
14.4 V
Battery status 80 %:
3 h 45 mins at 20 °C while
not in operation; battery status
100 %: approx. 5 h 40 mins
Automatic changeover to
trickle charging
Every 6 months
85 mins at -5 °C
85 mins at +21 °C
42 mins at +50 °C
Approx. 500 charging cycles in
approx. 4 years
Battery status indicator during operation and charging
-0.2 bar: 200 mins
-0.5 bar: 140 mins
-0.8 bar: 85 mins
ACCUVAC Lite
WM 11702
-5 °C to +50 °C
5 % to 95 % relative humidity
without condensation
540 hPa to 1100 hPa
ACCUVAC Lite
Lead
6.7 x 13.4 x 6.7 cm
1.15 kg
3.4 Ah
12 V nominal
Battery status 80 %:
2 h 40 mins
Battery status 100 %: 14 h
Automatic changeover to
trickle charging
Every 3 months
23 mins at -5 °C
40 mins at +21 °C
40 mins at +50 °C
400 charging cycles in approx.
3 years
-0.2 bar: 40 mins
-0.5 bar: 40 mins
-0.8 bar: 40 mins

Advertisement

Table of Contents
loading

This manual is also suitable for:

Accuvac liteWm 11602Wm 11702

Table of Contents