4 Preparation and operation
4.5.2 Placing the device in the wall mounting
Requirement
56
EN
ACCUVAC Pro/ACCUVAC Lite
2. Connect the vacuum tube and the connection hose for the
hygiene filter to the hygiene filter. When doing so, note:
The IN label on the hygiene filter must be facing the angled
connector for Serres secretions canister.
The vacuum tube and connection hose for the hygiene
filter must be pushed on as far as they will go.
3. Connect the connection hose for hygiene filter to the angled
connector for Serres secretions canister.
The wall mounting is installed as per the assembly instructions.
The 12 V connection cable or the power supply unit and
charger is clicked into place in the guide rail of the wall
mounting.
1. Connect the 12 V connection cable to a 12 V DC power source
or
Connect the power supply unit and charger to the line power.