Part
Cleaning
Secretions canister
Rinse off with clear
cover (upper sec-
water and clean
tion)
with a brush/cloth
Secretions canister
cover (lower sec-
tion)
Rinse off with clear
Float ball
water and clean
with a brush/cloth
Rinse out with
Reusable suction
warm, clear water
hose
for at least 10
seconds
Fingertip control
Disposable, do not reuse
Stages of machine cleaning and disinfection in a washer-disinfector according to ISO 15883-1:
Rinse for 1 minute with cold, demineralized water
Clean for 5 minutes at 55 °C (± 2 °C) with neodisher
Neutralize for 1 minute with demineralized water (1/3 cold water, 2/3 warm water)
Rinse for 1 minute with demineralized water (1/3 cold water, 2/3 warm water)
Thermally disinfect for 5 minutes at 93 °C with demineralized water
** Sterilization:* The service life of reusable elements is designed for a maximum of 50 reprocessing cycles.
Disinfection
Immersion
disinfection
(Recommendation:
GIGASEPT FF (new))
Rinse off with
clear water and
Immersion
clean with a
disinfection
brush/cloth
(Recommendation:
Clean the float
GIGASEPT FF (new))
ball guide with
a round brush
Immersion
disinfection
(Recommendation:
GIGASEPT FF (new))
Immersion
disinfection
(Recommendation:
GIGASEPT FF (new))
13. Allow the individual components of the reusable canister
system to dry.
Thermal
disinfection*
Disinfect at 93 °C in
a washer-disinfector
Disinfect at 93 °C in
a washer-disinfector
Disinfect at 93 °C in
a washer-disinfector
Disinfect at 93 °C in
a washer-disinfector
®
MediClean (0.5 % v/v)
ACCUVAC Pro/ACCUVAC Lite
5 Hygienic reprocessing
Sterilization**
If necessary, steam
sterilize at 134 °C,
3 x fractionated pre-
vacuum method,
sterilization time
5 mins with devices
according to EN 285
If necessary, steam
sterilize at 134 °C,
3 x fractionated pre-
vacuum method,
sterilization time
5 mins with devices
according to EN 285
If necessary, steam
sterilize at 134 °C,
3 x fractionated pre-
vacuum method,
sterilization time
5 mins with devices
according to EN 285
If necessary, steam
sterilize at 134 °C,
3 x fractionated pre-
vacuum method,
sterilization time
5 mins with devices
according to EN 285