Injection Valve; Metrohm Suppressor Module (Msm); Detector; Sample Degasser - Metrohm 930 Compact IC Flex ChS/Deg Manual

Table of Contents

Advertisement

■■■■■■■■■■■■■■■■■■■■■■
7.9

Injection valve

Actuator switch-
ing time
Maximum operat-
ing pressure
Material
7.10

Metrohm Suppressor Module (MSM)

Resistance to sol-
vents
Switching time
7.11

Detector

7.12

Sample degasser

Material
Resistance to sol-
vents
Time to establish
vacuum
7.13

Power connection

Required supply
voltage
Required fre-
quency
Power consump-
tion
Power supply unit
930 Compact IC Flex ChS/Deg (2.930.1260)
typ. 100 ms
35 MPa (350 bar)
PEEK
No restriction
typ. 100 ms
You can find the technical specifications for the detector in the manual for
the detector.
Fluoropolymer
No restriction (except PFC)
< 60 s
100 - 240 V ± 10% (autosensing)
50 - 60 Hz ± 3 Hz (autosensing)
65 W for typical analysis application
25 W standby (conductivity detector to 40 °C)
Up to 300 W maximum, electronically monitored
internal fuse 3.15 A
7 Technical specifications
■■■■■■■■