Metrohm 930 Compact IC Flex ChS/Deg Manual page 31

Table of Contents

Advertisement

■■■■■■■■■■■■■■■■■■■■■■
930 Compact IC Flex ChS/Deg (2.930.1260)
Adjust the length of the eluent aspiration tubing so that the aspi-
ration filter is at the bottom of the eluent bottle. Then fix it in
place using the M8 tubing nipple.
Seal the M6 opening on the bottle cap with the M6 threaded
stopper from the accessory set.
6 Mounting the adsorber tube
NOTE
Depending on the eluent used, the adsorber tube (6.1609.000)
must be filled differently:
For alkaline eluents or eluents with a low buffer capacity: first a
little cotton, then with CO
For all other eluents: only with cotton.
Remove the plastic cover from the large opening of the adsorber
tube. Fill the adsorber tube and close it again using the plastic
cover.
Insert the adsorber tube into the bottle cap's large opening. Fas-
ten it to the bottle cap using the SGJ clip (6.2023.020).
adsorber material.
2
3 Installation
■■■■■■■■
23