3 Pre-rinsing the aspiration filter
NOTE
Always wear gloves when handling the aspiration filter.
In order to avoid air bubbles after the installation of the aspiration fil-
ter, we recommend pre-rinsing the aspiration filter with water or elu-
ent.
For pre-rinsing, you will need the holder for the eluent aspiration fil-
ter (6.2744.360), a syringe and the purge needle (6.2816.040).
This procedure can also be found in the video "Inserting a new aspi-
ration filter" at the following link:
4 Mounting aspiration filter
NOTE
Always wear gloves when handling the aspiration filter.
Place the loose end of the eluent aspiration tubing into the aspira-
tion filter.
The end of the tubing should reach approximately to the center of
the aspiration filter.
Tighten the aspiration filter to the filter holder.
Figure 8
Installing tubing weighting and aspiration filter
5 Installing the eluent bottle cap on the eluent bottle
Insert the eluent aspiration tubing into the eluent bottle
(6.1608.070).
Tighten the bottle cap on the eluent bottle.
■■■■■■■■■■■■■■■■■■■■■■