5.8 Servicing the inline filter
5.8
Servicing the inline filter
Maintenance interval
Accessories
■■■■■■■■
70
Connecting the entry and exit for the high-pressure pump
1
Remove the stopper from the coupling. Tighten the coupling to
■
the pressure screw located on the pump head inlet capillary (1).
Reconnect the pump head outlet capillary to the pump head out-
■
let (2).
NOTE
You can find a video sequence for this task in the Multimedia Guide IC
Maintenance or on the Internet at http://ic-help.metrohm.com/.
The filter must be replaced at least every 3 months; it may need to be
replaced more frequently, depending on the application.
For this task, you need the following accessories:
Two adjustable wrenches (6.2621.000) from the accessory kit: Vario/
■
Flex Basic (6.5000.000)
Tweezers
■
A new filter from the packaging (6.2821.130)
■
■■■■■■■■■■■■■■■■■■■■■■
930 Compact IC Flex ChS/Deg (2.930.1260)