2
Operation and Features
Example 2, Set drive current at channel 1
& *CLS; *RST
& SOUR:VOLT:RANG R20V, (@1)
& SOUR:CURR:RANG R10mA,
& SOUR:VOLT:LIM 10V, (@1)
& SOUR:CURR 5mA, (@1)
& OUTP ON, (@1)
& MEAS:CURR? (@1)
$ 5.0E-03
40
(@1)
Resets the source to the default power-on state.
This command can be ignored if this operation is
not required.
Sets voltage range of channel 1 to 20 V.
Sets current range of channel 1 to 10 mA.
Sets voltage limit of channel 1 to 10 V.
Sets channel 1 to output 5 mA.
Turns on output for channel 1.
Queries the current for channel 1.