Example 2
This example stores the source current and measure voltage commands
into the active memory list (memory list 1) for channel 1 and executes
them when a remote trigger command is received.
To perform the function above, follow the sequence below.
& SOUR:MEM:LIST 1, (@1)
& SOUR:MEM:LIST:CLE, (@1)
& SOUR:MEM:VOLT:RANG R20V,
& SOUR:MEM:CURR:RANG
& SOUR:MEM:VOLT:LIM 20,
& SOUR:MEM:CURR:SOUR 0.12,
U2722A/U2723A User's Guide
a Select an active memory list.
b Clear the content of the active memory list.
c Store the "set voltage range" command into the active memory list.
d Store the "set current range" command into the active memory list.
e Store the "set voltage limit" command into the active memory list.
f Store the "drive current" command into the active memory list.
g Store the "set output on" command into the active memory list.
h Store the "measure voltage" command into the active memory list.
i Trigger the execution of commands in the active memory list.
(@1)
R120mA, (@1)
(@1)
(@1)
Operation and Features
This command selects memory list 1 as the
active memory list for channel 1. This command
can be omitted if the default active memory list
has not been changed.
This command clears all the content in the
active memory list of channel 1.
This command stores the "set voltage range to
20 V" command into the active memory list of
channel 1.
This command stores the "set current range to
120 mA" command into the active memory list of
channel 1.
This command stores the "set voltage limit to
20 V" command into the active memory list of
channel 1.
This command stores the "drive current to
120 mA" command into the active memory list of
channel 1.