Agilent Technologies U2722A User Manual page 83

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

3
Characteristics and Specifications
U2723A USB Modular Source Measure Unit performance specifications
Voltage programming
12 months (at 23 ° C ± 5° C), ± (% of output + offset)
Current programming
12 months (at 23° C ± 5 ° C), ± (% of output + offset)
Voltage readback
12 months (over USB with respect to the actual
output at 23 ° C ± 5 ° C), ± (% of output + offset)
Current readback
12 months (over USB with respect to the actual
output at 23 ° C ± 5° C), ± (% of output + offset)
[1] Accuracy measurements are based on NPLC 10.
54
Range
Accuracy
± 2 V
0.075% + 1.5 mV
± 20 V
0.05% + 10 mV
± 1 μA
0.085% + 0.85 nA
± 10 μA
0.085% + 8.5 nA
± 100 μA
0.075% + 75 nA
± 1 mA
0.075% + 750 nA
± 10 mA
0.075% + 7.5 μA
± 120 mA
0.1% + 100 μA
± 2 V
0.075% + 1.5 mV
± 20 V
0.05% + 10 mV
± 1 μA
0.085% + 0.85 nA
± 10 μA
0.085% + 8.5 nA
± 100 μA
0.075% + 75 nA
± 1 mA
0.075% + 750 nA
± 10 mA
0.075% + 7.5 μA
± 120 mA
0.1% + 100 μA
U2722A/U2723A User's Guide
[1]
Resolution
0.1 mV
1 mV
100 pA
1 nA
10 nA
100 nA
1 μA
20 μA
0.1 mV
1 mV
100 pA
1 nA
10 nA
100 nA
1 μA
20 μA

Advertisement