Operation and Features
For example, to set the time interval to 40 ms per measurement with 1500
samples, use the following command.
SENS:SWE:TINT 40, (@1); :SENS:SWE:POIN 1500, (@1)
Only the MEASure:ARRay:VOLTage[:DC]? (@1|2|3)and
MEASure:ARRay:CURRent[:DC]? (@1|2|3)commands are governed by
the number of data points and time interval settings.
N O T E
46
You are required to configure the time-out settings via the Agilent IO
Libraries when different values of the sweep points and time interval are
set. For more information on how to set the time-out and what are the
recommended time-out settings, refer to
Settings"
on page 60.
"Appendix B: Time-out