Collimator, X-Iris

NOTE
WARNING
Attach the center cross in the center of the image intensifier input screen.
Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
Start the service program.
Select the "Adjustment, Calibration" menu.
Answer the query with "yes".
Select "X-Iris" in the "Function Groups" combo box.
In the window "Available Functions:" select "1. Init (Limits Off / Autosearch)" and click on
Execute.
Close the status window by clicking on the OK button.
For the query "Do you want to reset the Unit", click "No".
The X-Iris can now be completely opened and closed with the X-Iris keys.
Select I.I. survey format.
During fluoroscopy, adjust the X-Iris by opening and closing it with the X-Iris keys so that
the collimator blades are still visible along the edge of the I.I.
Select "2. II. - Full Format Position" on the service PC and click on "Execute" .
Close the status window by clicking on the OK button.
For the query "Do you want to reset the Unit", click "No".
Switch to zoom format.
During fluoroscopy, adjust the X-Iris by opening and closing the X-Iris keys so that the
collimator blades are still visible along the edge of the I.I.
Select "3. II. - Zoom Format Position" on the service PC and click on "Execute" .
Close the status window by clicking on the OK button.
For the query "Do you want to reset the Unit", click "No".
Open the X-Iris completely.
Select "4. CEX Position" on the service PC and click on "Execute" .
Close the status window by clicking on the OK button.
For the query "Do you want to reset the Unit", click "No" .
Close the X-Iris completely.
Release fluoroscopy. The X-Iris must close to < 4.5 cm .
Siemens AG
Medical Solutions
Because of the mechanical hysteresis of the semi-transparent
slot diaphragm, open the collimator blades to maximum aperture,
prior to adjusting the radiation iris.
X-ray radiation!
See Chapter 1, Safety Information.
Observe the need for radiation protection. Wear a lead apron.
SPR2-230.061.01
Rev. 13
Page 11 of 30
12.04
CS PS 24
5 - 11
5
SIREMOBIL Iso-C
System Manual