Connecting The Power Cable; Connecting The Monitor Trolley Cable; Inserting The On/Off Assembly; Connecting The Ground Wire - Siemens ARCADIS Varic Replacement Instructions Manual

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

Installing the ON/OFF assembly

Connecting the power cable

Pull the anti-kink sleeve of the power cable (1/Fig. 1) over the power cable so that
approx. 13 cm of the gray insulated cable are exposed. See Fig. 1.
Run the power cable through the cable opening to the cable clamp (Fig. 2). The groove
of the anti-kink sleeve (2/Fig. 1) rests in the sheet metal notch (2/Fig. 2) of the ON/OFF
assembly.
Reattach the power cable with the cable clamp.
Connect the power cable to the power terminal:
X1. ground wire (yellow/green terminal - yellow/green cable)
X1.N (blue terminal - blue cable)
X1. L1 (gray terminal - brown cable)

Connecting the monitor trolley cable

Run the monitor trolley cable through the cable opening to the two cable clamps (Fig. 2).
Attach the gray sheath of the monitor trolley cable (strain relief) with the clamp 3/Fig. 2.
Connect the shield of the monitor trolley cable to ground using the clamp 4/Fig. 2.
In doing so, ensure good ground contact of the shield.

Inserting the ON/OFF assembly

Lift the ON/OFF assembly into the monitor trolley by its rear side and push the assembly in.
For better accessibility, the ON/OFF assembly should project out of the monitor trolley
approx. 10 cm.

Connecting the ground wire

Connect the ground wire coming from the monitor trolley together with the ground wires
from the monitor trolley cable to the ground wire terminal (base plate of the ON/OFF
assembly) and tighten them with screws (5/Fig. 2).
In doing so, ensure good contact of the ground wire with the ON/OFF assembly.

Wiring of D66.X1 and D66.X5

Run cables D66.X1 and D66.X5 through the existing cable duct of the monitor trolley to
the PC (imaging system) again, plug them into D66.X1 and D66.X5 and fasten them.

Wiring of CAN cable (3D Reconstruction option)

If the 3D Reconstruction option is present, run the CAN connector to the CAN-to-USB
converter again, plug it in and fasten it.
Siemens AG
Medical Solutions
n.a.
SP00-000.841.01
Rev. 01
Page 3 of 10
08.05
CS PS 24
3 - 3
3
3
3
3
3
3
ARCADIS
System Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Arcadis orbic

Table of Contents