Multispot 2000; Area Dose Product Meter; Laser Targeting Device - Siemens SIREMOBIL Iso-C Service Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:
Table of Contents

Advertisement

6 - 34

MULTISPOT 2000

Before installing the new MULTISPOT 2000, set the correct refresh rate. Refer to Table 1.
Jumper
X10
X11
Tab. 1
Disconnect the SIREMOBIL line voltage plug from the receptacle.
Replace the MULTISPOT 2000.
The screws securing the MULTISPOT 2000 are accessible on the underside of the cam-
era.
Perform a camera function test after the replacement.
Generate and program the exposure parameters for the Multispot 2000 according to the
IQ Quick test (refer to the log book). Record the new exposure parameters in the IQ
Quick test instructions.

Area dose product meter

Remove the SIREPHOS cover.
Replace the area/dose product measurement device.
Perform the calibration of the area dose product measurement device. Refer to chapter 5
"Calibrating the Area Dose Product Measurement".
Test the accuracy of the area dose product measurement device. Refer to chapter 5
"Testing the accuracy of the area dose product measurement device.
Refit the SIREPHOS cover.
Seal the SIREPHOS cover using sealing compound. See Chapter 1, "Covers".

Laser Targeting Device

Remove the SIREPHOS cover.
Replace the laser diodes after removing the plastic bracket and desoldering the wires.
Adjust the laser diodes using the adjustment device supplied with the laser targeting de-
vice.
Refit the SIREPHOS cover.
Seal the SIREPHOS cover using sealing compound. See Chapter 1, "Covers".
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
50 Hz Refresh rate
Position
X4
b
X5
open
X6
b
X7
b
X8
b
X9
b
b
b
60 Hz Refresh rate
Jumper
X4
X5
X6
X7
X8
X9
X10
X11
SPR2-230.061.01
Rev. 13
12.04
Position
a
open
a
a
a
a
a
a
Page 34 of 48
CS PS 24
6
6
6
Siemens AG
Medical Solutions