Baumer
11.10.11
LinePWMMode
Selects the PWM mode of the selected Output line.
Name
Category
Interface
Access
Unit
Values
Time diagrams of PWMModes:
Fixed Frequency
Trigger
Exposure
LineSource e.g.
ExposureActive
Signal Output
LinePWMDuration
LinePWMMaxDuration
11.10.12
LinePWMOffTime
Switch-off time in microseconds included in the PWM period.
Name
Category
Interface
Access
Unit
Values
V1.1 | VCXG.2 / VCXG.2.XC / VCXG.2.I / VCXU.2
LinePWMMode
DigitalIOControl
IEnumeration
Read / Write
-
Fixed Frequency
The selected Output generates a permanent pulse fre-
quency beginning with every transition from 0 to 1 and end-
ing with every transition from 1 to 0.
Off
PWMMode is off. Output reaction is the same as of stan-
dard Output.
OnePulse
The selected Output generates one pulse with every transi-
tion from 0 to 1.
Off
Exposure
LinePWMOffTime
LineSource e.g.
ExposureActive
Signal Output
LinePWMPeriodTime
LinePWMOffTime
DigitalIOControl
IInteger
Read only
µs
PWM setting -specific.
Trigger
Operating Manual
GenICam Camera Features
OnePulse
Trigger
Exposure
LineSource e.g.
ExposureActive
LinePWMOffTime
Signal Output
LinePWMDuration
LinePWMMaxDuration
LinePWMPeriodTime
|
11