Checking / Replacing The Particle Filter - TESTO 330 Instruction Manual

Flue gas analyzer
Hide thumbs Also See for 330:
Table of Contents

Advertisement

1. Unlock the condensate trap and pull it vertically off the meas.
box.
2. Open the condensate outlet on the condensate trap: Pull out
approx. 5 mm against the stop.
3. Let the condensate run out into a sink.
4. Wipe off any drops still on the condensate outlet with a cloth
and close the condensate outlet.
Pos: 86 /TD/Produkt instand halten/testo 330/testo 330 Instandhaltung Partikelfilter prüfen @ 6\mod_1279265518333_7011.doc @ 66965 @
7.9.

Checking / replacing the particle filter

Checking the particle filter:
Check the particle filter of the modular flue gas probe periodically
for contamination: check visually by looking through the window
of the filter chamber.
Replace the filter if there are signs of contamination.
Replacing the particle filter:
The condensate outlet must be completely closed
(marking), otherwise measuring errors could be caused by
infiltrated air.
The filter chamber may contain condensate.
7 Maintaining the product
65