Checking / Replacing The Particle Filter; Basic Operating Steps; Switching The Measuring Instrument On - TESTO 330-1 Instruction Manual

Flue gas analyzer
Table of Contents

Advertisement

E. Operation
20
E.3 Regular care
E.3.2 Checking / replacing the particle filter
Checking the particle filter:
Replacing the particle filter:
E.4 Basic operating steps
E.4.1 Switching the measuring instrument on
.
- The start screen is displayed (for about 5 s).
- Display illumination is switched on for 10 s.
Option:
To go directly to a measurement while the start screen is being displayed, press
the function key for the desired measurement. See also Start key configuration ,
p. 29.
- The
Measurements
-o o r-
- If another probe / sensor is connected rather than a flue gas probe: the measuring
menu for that probe / sensor is opened.
Particle filters of the modular flue gas probe must
be checked regularly for contamination:
Check visually by looking through the window of
the filter chamber.
Replace the filter if there are signs of
contamination.
Filter chamber may contain condensate.
1 Open the filter chamber: Turn gently
anticlockwise.
2 Remove the filter plate and replace it with a
new one (0554 3385).
3 Fit the filter chamber and lock it: Turn gently clock-
wise.
menu is opened.