7.8.

Replacing the probe module

✓ Disconnect the flue gas probe from the measuring instrument.
1. Press the button on the top of the probe handle (1) and remove
the probe module (2).
2. Plug in the new probe module and lock it in place (3).
7.9.

Checking/replacing the particle filter

Checking the particle filter:
Check the particle filter of the modular flue gas probe periodically
for contamination: check visually by looking through the window
of the filter chamber.
> If there is visible contamination or inadequate pump flow,
replace the filter.
Replacing the particle filter:
The filter chamber may contain condensate. This is not a
malfunction and will not cause wrong readings.