5 First steps
5.5.

Instrument diagnosis

Important operating values and instrument data are displayed. A
main test can be carried out. Instrument errors that have not yet
been rectified can be displayed.
Call up function:
>
[
Error diagnosis
>
Error diagnosis
-
Unresolved errors, warnings and notes are displayed.
> View next/previous error: [▲], [▼].
Instrument information
>
Device Information
-
Information is displayed.
Tightness test
1.
Main test
2. Short both gas connections through the connection hose.
3. Start the check: [OK].
-
Zeroing of pressure sensor
-
Pressure is built up and measurement is carried out
-
The result of the measurement is indicated by a traffic light.
Gas feeding unit test
1.
Gas feeding unit test
2. Connect the measuring instrument to the gas feeding unit.
3. Start the test: [OK].
-
If the gas feeding unit is full, it is emptied.
4. Gas feeding unit is filled.
-
The measurement result is indicated by a traffic light.
5. Gas feeding unit is emptied.
6. Exit testing:
30
]
Instrument diagnosis
→ [OK].
→ [OK].
Tightness test
Settling period expires (15 min)
Measuring time expires (5 min)
[Back]
→ [OK].
[OK]
→ [OK].