E.6 Instrument diagnosis

Important operating values and device data are displayed. A gas path check
(testo 330-2, -3 / testo 330-2 LL) can be carried out. The status of the measuring cells
and device errors which have not yet been rectified can be viewed.
Calling up the function:
Inst' diagnosis
-o o r-
.
Carrying out a gas path check (testo 330-2, -3 / testo 330-2 LL)
1
Gas path check
2 Place the black sealing cap on the tip of the flue gas probe.
- The pump flow is displayed. If the flow rate is ≤0,02 l/min, the gas paths are not leak-
ing.