Applicable Input Modules
• 8967 TEMP Unit
Connect to the terminal block on the input
module.
Insert to terminal block
25 mm
Outer Insulation
Inner insulation
Inserting a Thermocouple
Connection Holes
5
Attach to the measurement object
Measuring Temperature
Use to connect: Thermocouple
(Compatible wire: AWG 16 to 26, 0.4
to 1.2 mm diameter)
10 mm
1
Thermocouple
single wires
Terminal Block
3
2
4
If surrounding equipment is affected by noise, coil the ther-
mocouple several times and then attach the included ferrite
clamp-on choke (as seen in the diagram to the right).
Required item:
Thermocouple, Ferrite clamp-on choke (8967's option),
flat-blade screwdriver (2.6-mm blade)
Recommended wire:
Compatible wire:Single-strand thermocouple wire, 0.4
to 1.2-mm diameter
1
Strip insulation from the thermocouple
wires as shown at the left.
Stripping length: approx. 10 mm
2
Push the blade of a flat screwdriver
into the button on the terminal block of
the input module.
3
Insert each thermocouple wire into the
appropriate terminal hole while press-
ing the button.
Confirm proper polarity.
4
Release the button.
The thermocouple is connected.
5
Attach to the measurement object.
To remove the thermocouple
Hold the button while pulling the ther-
mocouple wire out.
25
2.2 Connecting Cords
Connect to terminal block
2
Need help?