9 Settings

9.2 Program settings

9.2.1 Programs (2100)

9.2.1.1 Turning programs ON or OFF
About this task
Steps
9.2.1.2 Permanently adjusting preset exposure values
About this task
Steps
62
KaVo ProXam
Follow these steps to turn a program(s) ON or OFF.
1. Select Program > 2100 Programs.
2. Select a program group (e.g. 2D Panoramic).
3. Turn a program type(s) (e.g. Interproximal) ON or OFF.
4. Select the green check mark button.
NOTE
The function is available for programs that have been activated in menu
2300 Licences.
Follow these steps to permanently adjust preset exposure values.
1. Select Program > 2100 Programs.
2. Select a program group (e.g. 2110 2D Panoramic).
3. Select a program type (e.g. Interproximal).
4. Select the exposure values you wish to adjust (e.g.70kV / 10 mA for
patient size M).
In 2D panoramic programs select also the MultiView button if you
wish to adjust the presets for the MultiView imaging mode.
In 3D programs the exposure values are given separately for each
image resolution. The image resolutions that are not available are
shown with faded buttons. Select also the ULD (Ultra Low Dose)
button if you wish to adjust the presets for the ULD function.
5. Use the minus or plus buttons to set the exposure values you wish to
use.
6. Select the green check mark button.
7. Repeat for an other program type, patient size or image resolution (3D) if
needed.