8.3 Selecting exposure settings

Instructions for use
1. Select the patient.
2. Right-click on the patient and select Capture followed by 3D Capture.
Romexis user's manual for details on Romexis functions.
Refer to the
Ask the patient to remove any spectacles, hearing aids, dentures, hairpins,
and personal jewellery such as earrings, necklaces and piercings as these
can produce shadows or reflections in the image. The patient should also
remove any loose items of clothing (e.g. scarf, tie) that might get caught in
the arm structures of the X-ray unit.
NOTE
High contrast objects, such as gold teeth or amalgam, may cause artefacts
in the image.
Place a protective lead apron over the patient's back if required.
Refer to section "Touch screen" on page 15 for general information on how
to make or cancel selections on the touch screen.
8 3D patient exposure
KaVo ProXam
33