Instructions for use

Advertisement

Table of Contents
loading

Summary of Contents for KaVo ProXam iS

  • Page 1 KaVo ProXam iS Instructions for use...
  • Page 2 The manufacturer, assembler and importer are responsible for the safety, reliability and performance of the unit only if: - installation, calibration, modification and repairs are carried out by qualified authorised personnel - electrical installations are carried out according to the appropriate requirements such as IEC 60364 - equipment is used according to the operating instructions.
  • Page 3: Table Of Contents

    Supported operating systems.......................21 14.5 Operating environment......................... 21 14.6 Transportation and storage environment..................21 14.7 Distributor and manufacturer......................21 Appendix A: Exposure value tables for KaVo ProXam iX................... 23 Default exposure values....................... 23 Preprogrammed settings values....................23 Instructions for use KaVo ProXam iS...
  • Page 4 Table of contents KaVo ProXam iS Instructions for use...
  • Page 5: Introduction

    1 Introduction 1 Introduction This Instructions for use describes how to use and install the KaVo ProXam iS sensor that is intended to be used for capturing digital intraoral X-ray images from patient’s jaw, teeth, gums, roots and root canals by trained dental care professionals.
  • Page 6: Symbols

    Planmeca ProSensor is ETL classified and conforms to ANSI/AAMI ETL CLASSIFIED ES60601-1 and is certified to CAN/CSA C22.2 No. 60601.1. 3143029 Separate collection for electrical and electronic equipment according to Directive 2002/96/EC (WEEE) KaVo ProXam iS Instructions for use...
  • Page 7: Safety Precautions

    Never cut, nick or sharply bend the sensor cable. Always advise the patient not to bite the sensor or the cable. The KaVo limited warranty does not cover damage which is due to misuse, e.g. dropping the sensor, neglect, or any cause other than ordinary application.
  • Page 8: Reporting Serious Incidents

    4 Safety precautions 4.1 Reporting serious incidents Serious incidents that have occurred in relation to the device must be reported to the manufacturer and the local competent authority. KaVo ProXam iS Instructions for use...
  • Page 9: Control Box Indicator Light

    Purple light at beginning when powering up the Control box is running a factory software control box NOTE The exposure can only be taken when the control box indicator light is green and steady, not when the indicator light is flashing. Instructions for use KaVo ProXam iS...
  • Page 10: Before Exposure

    6 Before exposure 6 Before exposure NOTE Detailed instructions for using KaVo ProXam iX X-ray unit and Romexis software are given in their Instructions for use, which should be used in conjunction with this Instructions for use. NOTE It is recommended to use a sensor holder. Select the correct sensor holder according to the type of exposure, refer to the sensor holder Instructions for use supplied with the sensor holder package.
  • Page 11 6 Before exposure Paralleling technique (recommended) Place the sensor (2) to a sensor holder and align the holder parallel to the long axis of the tooth (1). Use a long cone for the paralleling technique. Instructions for use KaVo ProXam iS...
  • Page 12: Selecting Exposure Values

    On how to place the sensor into the patient’s mouth refer to the sensor holder manual supplied with the digital sensor. 6.3 Selecting exposure values The following tables show the recommended exposure values for KaVo ProXam iS. Short cone 20 cm (8”) exposure values...
  • Page 13 MAND MAXI 66 kV/ 8 mA adult MAND MAXI 63 kV/ 8 mA adult MAND MAXI 60 kV/ 8 mA adult MAND I = INCISORS, M = MOLARS, P = PREMOLARS AND CANINES Instructions for use KaVo ProXam iS...
  • Page 14: Capturing Intraoral Images

    After the exposure the message Saving the image appears on the display and the image is automatically stored into the database. 4. Define the tooth numbers and sensor orientation. The sensor orientation is indicated in the image by a grey triangle. KaVo ProXam iS Instructions for use...
  • Page 15: Capturing Images Into Study Template

    2. Select the desired study template from the list. At the beginning of the list there are empty templates and at the bottom of the list there are studies with dates that already include images captured earlier for the selected patient. Instructions for use KaVo ProXam iS...
  • Page 16 Exposure appears on top of the window. You can now take exposures as usual. Saving the image appears on the After the exposure the message display and the image is automatically stored into the database. KaVo ProXam iS Instructions for use...
  • Page 17 To cancel the process click Cancel. The captured images are saved and the incomplete study is preserved for later use. 6. Once all images have been captured click Done. Instructions for use KaVo ProXam iS...
  • Page 18: Image Quality Control

    Quart phantom or similar. It is recommended to regularly monitor the image quality using the same phantom according to the requirements of local authorities. See also KaVo ProXam X-ray unit intraoral systems Constancy test Instructions for use.
  • Page 19: Cleaning And Disinfection

    Use a new disposable hygienic sheath for every sensor usage. NOTE The sensor connector can be cleaned using a soft cloth. 10.2 Sensor holders For cleaning the sensor holders refer to the manual supplied with the sensor holder package. Instructions for use KaVo ProXam iS...
  • Page 20: Control Box

    Switch off the unit before cleaning. NOTE Do not disinfect the unit. CAUTION Never detach the control box ethernet cable without releasing the latch on the cable connector. Forcefully detaching the cable will damage the control box. KaVo ProXam iS Instructions for use...
  • Page 21: Service

    11 Service All cyber security software updates listed in a technical bulletin must be installed on the X-ray unit. KaVo Technical Service If you have any questions or complaints, please contact the KaVo Technical Service: +49 (0) 7351 56-2900 service.xray@kavo.com...
  • Page 22: Terms And Conditions Of Warranty

    Claims from this warranty can only be asserted when the transfer form (copy) belonging to the product has been sent to KaVo, and the original can be presented by the operator or user. Exceptions to the standard warranty •...
  • Page 23: Disposal

    TPE/PU Packing cardboard, paper, PE foam Sensors Return the sensors to KaVo. Other parts NOTE If the component boards cannot be recycled handle them as electronic scrap, i.e. according to the local legislation. Instructions for use KaVo ProXam iS...
  • Page 24: Technical Specification

    112 x 46 x 24 mm (4.41 x 1.81 x 0.94 in.) Power input 48 V DC 65 mA Cables ControlBox to PoE RJ45 10m OR 15m PoE to LAN RJ45 10m OR 15m KaVo ProXam iS Instructions for use...
  • Page 25: Usb Controlbox

    95%. Atmospheric pressure range should be between 700 hPa - 1060 hPa. 14.7 Distributor and manufacturer Distributor KaVo Dental GmbH, Bismarckring 39, 88400 Biberach, Germany Phone +49 7351 56-0, Fax +49 7351 56-1488 http://www.kavo.com Instructions for use KaVo ProXam iS...
  • Page 26 14 Technical specification Manufacturer Planmeca Oy, Asentajankatu 6, FIN-00880, Helsinki, Finland Phone: +358 20 7795 500, Fax: +358 20 7795 555, http:// www.planmeca.com KaVo ProXam iS Instructions for use...
  • Page 27: Appendix A: Exposure Value Tables For Kavo Proxam Ix

    Appendix A: Exposure value tables for KaVo ProXam iX Appendix A: Exposure value tables for KaVo ProXam A.1 Default exposure values The following table shows the default exposure values for KaVo ProXam iS with no target selected. Short cone Long cone...
  • Page 28 Appendix A: Exposure value tables for KaVo ProXam iX OCCLUSAL ENDODONTIC BITE-WING time time time Maxilla Child 0,125 0.125 Mandible KaVo ProXam iS Instructions for use...