10 Help messages

10 Help messages
Code
H101
Exposure switch
H102
H103
ProTouch desktop
application
H105
Emergency stop
button
H110
Device
Maintenance
H112
Collimator
H113
H120
Subscription
H121
H130
Patient safety area
H142
Height movement
68
KaVo ProXam
The X-ray unit incorporates a self-checking feature that monitors the
operation of the unit. If the system detects an operating error, a help
message (e.g. H101) appears on the touch screen.
The X-ray unit will not accept any commands from the user until the help
message is cleared from the touch screen. Clear the message by touching
the green check mark.
The following list shows, in numerical order, all the help messages that can
appear.
Explanation
The exposure button was released
before the end of the exposure.
The exposure button is stuck or the
cable is short circuited.
The ProTouch desktop application
needs to be closed.
The emergency stop button has
been activated.
The date for your next scheduled
device maintenance is due.
Check panoramic X-collimator
position setting.
Check panoramic X-collimator width
setting.
A subscription licence is expiring
soon.
A subscription licence has expired.
Patient safety area issue detected.
Height movement is not possible
because one (or more) of the
positioning control buttons or the
positioning joystick is stuck.
Comments
Guide the patient away from
the X-ray unit before moving
the C-arm.
Press and hold down the
exposure button for the entire
duration of the exposure.
Release the exposure button.
Contact your service
technician if you need to
replace the exposure switch.
Restart the application.
All movements of the X-ray
unit are blocked and no
radiation is generated.
Guide the patient away from
the X-ray unit. Then release
the emergency stop button to
resume normal operation.
Please contact your service
technician to arrange a
service visit.
Check collimator settings.
Check collimator settings.
Please renew the
subscription to avoid losing
features.
Some features have been
disabled.
Check the scan and layer
settings.
Clear any obstruction before
moving the column again.
Instructions for use