6.10.2 Positioning joystick

6.10.3 Image volume up / down

22
KaVo ProXam
NOTE
The column movement stops automatically if the emergency stop plate at the
bottom is pressed upwards. Clear any obstruction before moving the column
again.
NOTE
When positioning seated patients (for example, in a wheelchair) always first
move the X-ray unit down before you position the patient in the unit.
Use the image volume up and down buttons for vertical adjustment of the
image volume in 3D imaging. Use the buttons when the patient is positioned
in the X-ray unit.
The X-ray unit moves slowly when you press the up / down button. The
height of the patient support table is not changed, i.e. the patient remains in
the same position.