KaVo ProXam Extraoral
installation quick guide

Advertisement

loading

Summary of Contents for KaVo ProXam Extraoral

  • Page 1 KaVo ProXam Extraoral installation quick guide...
  • Page 2: Tools Required

    ProXam 3DQ 1170 1000 46.0” 39,4” (50) (50) 13,8” 37,4” (2,0”) (2,0”) Unit dimensions = 1300 / 51,2” Unit dimensions = 2120 / 83,5” Minimum recommended space = 2220 / 87,4” TOOLS REQUIRED MOUNTING ACCESSORIES 1. Tape measure 2. Flat head screwdriver 3.
  • Page 3 UNPACKING DELIVERY CONTENTS (WOODEN SHIPPING CRATE) (CARDBOARD SHIPPING CRATE) Package 1 Package 2 (3D system) 1. KaVo ProXam X-ray unit 1. Ethernet switch 2. Stationary column side covers 2. 3D sensor 3. Shoulder arm cover 3. Reconstruction PC (RecoPC) 4. Accessories box...
  • Page 4 LIFTING UP X-UNIT ATTACHING X-RAY UNIT TO WALL AND FLOOR Wall made of concrete or stone Wooden wall 1250 mm 3310 N (338 kg; 744 lb) NOTE! The wall support installation must be able to resist a pull-out force of 3310 N (338 kg;...
  • Page 5 ATTACHING X-RAY UNIT TO WALL WITH SECOND ADJUSTABLE WALL SUPPORT Wall made of concrete or stone Wooden wall 180° 1250 mm 3310 N (338 kg; 744 lb) 125mm NOTE! The wall support installation must be able to resist a pull-out force of 3310 N (338 kg;...
  • Page 6 UNSCREW THE TRANSORT LOCKING BAR REMOVE THE TRANSPORT LOCKING BAR REMOVE THE SUPPORT FRAME AND STYROFOAM PACKAGING CONNECTING MAINS CABLE X-RAY UNITS WITH DETACHABLE POWER SUPPLY CORD Attach strain relief to the junction box Remove a knockout in the juction box. Junction box Mains X-ray unit power cable...
  • Page 7 REMOVING TRANSPORT LOCKING INSTALLING MOVABLE PIN (90kV ROTATING TUBE HEAD) EXPOSURE SWITCH INSTALLING STATIONARY CONNECTING CABLES ON 2D CONNECTING EXPOSURE EXPOSURE SWITCH SWITCH CABLE SYSTEMS Romexis PC KaVo_ProXam_IP6.eps...
  • Page 8 CONNECTING CABLES ON 3D SYSTEMS Reconstruction PC Romexis PC Ethernet switch REMOVING ESD TAPE ATTACHING COVERS ATTACHING SENSOR HEAD KaVo_ProXam_IP7.eps...
  • Page 9: Post-Installation Checklist

    Face 3D calibration 3D beam check..................3D flat field calibration................3D geometry calibration................. 3D quality assurance (Q/A) test............. Other: KaVo ProXam Cephalostat calibration _______________________________________________________ _______________________________________________________ (If the X-ray unit is equipped with cephalostat) _______________________________________________________ Beam check................... _______________________________________________________ Flat field calibration................