Factory presets for image resolution High Definition (HD)
Child (XS)
Small adult (S)
Medium-sized adult (M)
Large adult (L)
Extra large adult (XL)
Factory presets for image resolutions Hi Res and Endo
Child (XS)
Small adult (S)
Medium-sized adult (M)
Large adult (L)
Extra large adult (XL)
8.6 Selecting patient movement correction
Steps
Instructions for use
Patient size
Patient size
If you need to adjust the preset exposure values for this exposure:
1. Select the kV / mA field.
2. Use the minus or plus buttons to set the exposure values you wish to
use. To improve the image contrast, reduce the kV value. To reduce the
radiation dose, reduce the mA value.
3. Select the green check mark button.
NOTE
You can adjust the preset exposure values permanently as described in
section "Programs (2100)" on page 62.
1. Use this button to select the CALM patient movement correction
function.
This is an algorithm that detects patient movement during exposure
and then compensates for the effects of the movement during image
reconstruction.
kV value
mA value
90
5
90
6.3
90
8
90
10
90
11
kV value
mA value
90
6.3
90
8
90
10
90
11
90
11
8 3D patient exposure
mA value with
ULD
4.5
5.6
7.1
9
11
mA value with
ULD
5.6
7.1
9
10
11